AATCC-Logo_web copy.jpg

When do AATCC scales REALLY need to be replaced?

Posted October 21, 2021

 

The scales used by textile testing labs are carefully designed and checked for consistency, but they can change with time, exposure and handling. There are some practical care and replacement guidelines for ensuring repeatable evaluations using AATCC scales.

 

Gray Scales

 

AATCC EP1 Gray Scale for Color Change and AATCC EP2, Gray Scale for Staining include an Appendix specifying the color difference for each chip pair and how to verify it.

 

Color difference for each pair of chips may be measured quarterly on a spectrophotometer with sphere geometry. Use a small area view (approximately 10-mm diameter) and specular component included. Be careful to position the scale so that only the desired chip is in the viewing area for each measurement. If ΔECIELAB is within the stated tolerance for all pairs, the gray scale is still suitable for use.

 

If you do not have access to a spectrophotometer or choose not to measure your scales every three months, replace each gray scale within one year of its initial use. For your convenience, there is a line on the scale to record the date of initial use. Of course, if there is visible damage to any of the chips or the mask, the scale should be replaced immediately.

 

Other recommendations for use and care are also listed on the scales and in the evaluation procedures.

 

  • Store scale in sleeve when not in use.

  • Do not place mask inside folded scale.

  • Inspect frequently for fingerprints and any other marks.

  • Replace annually or measure quarterly and replace as needed.

 

Replace your AATCC Gray Scale for Color Change here.

Replace your AATCC Gray Scale for Staining here.

 

AATCC 9-Step Chromatic Transference Scale

 

Guidelines for the Chromatic Transference Scale are similar to those for the Gray Scales. AATCC EP8 states, “The cleanliness and physical condition of the Scale is extremely important in obtaining consistent results.”

 

  • Inspect frequently for fingerprints and any other marks.

  • Inspect for physical damage to the chips (e.g., chipped edges, loose or buckled chips).

  • Replace or measure at least annually on a spectrophotometer.

 

Spectrophotometer measurements can be particularly challenging on the chromatic scale because of the small chip size. The measurement area should not include any open area or adjacent chip. The viewing port diameter must be smaller than 10 mm.

 

Replace your AATCC 9-step Chromatic Transference Scale here.

 

Stain Release Scales

 

Stain scales for AATCC TM175, Stain Resistance: Pile Floor Coverings and AATCC TM130, Soil Release: Oily Stain Release Method should be replaced annually – more often if they are visibly soiled or damaged. To avoid premature color change, it is best to store both scales in a dark place between uses.

 

  • Inspect frequently for fingerprints and any other marks.

  • Store away from light.

  • Replace annually.


Replace your AATCC Red 40 Stain Scale (for carpet) here.

Replace your Stain Release Scale here.

 

3D Appearance Replicas

 

The scales for appearance (wrinkle recovery, fabric smoothness, crease) evaluation are made up of three-dimension plastic replicas cast from fabric. While these are more durable than the printed scales, they still require careful handling and regular replacement.

 

  • Wipe with a damp cloth and allow to dry

  • Inspect frequently for marks, cracks, or other damage

  • Store replicas carefully in box slots.

  • Replace every three years.


Replace your AATCC 3D Smoothness Appearance Replicas here.

Replace your AATCC 3D Crease Replicas here.

Replace your AATCC 3D Wrinkle Recovery Replicas here.

 

Photographic Scales

 

Several AATCC test methods use photographic scales for evaluation. Some of these scales are mounted on rigid cards, while others are simply a sheet of paper.

  • Inspect frequently for marks, tears, scratches, or other damage

  • Replace every three years.

 

Replace your AATCC Photographic Spray Test Evaluation Scale here.

Replace your AATCC Photographic Seam Smoothness Scale here.

Replace your AATCC Photographic Flock Scale here.

Replace your AATCC Photographic Filter Residue Scale here.

Replace your Photographic Dye Dusting Scale here.

 

Source: AATCC

  • LinkedIn
  • Wix Facebook page
  • Instagram
  • Wix Twitter page
STA_ad_011520-optimized_compressed.gif
ATMA_Techtextil 2021 ad_website_011121-O
Lang_Ligon_eTC_inside_pages_ad_2021-comp
AB Carter-website ad_051021_DRAFT1.gif